Density meters for manual sample supply Set 1 / Set 2

Density meter DS7800 with syringe for manual sample supply

to read more details

The right decision for low sample volumes

For manual sample delivery via syringe, we offer two different density meter sets: For low sample volumes and high viscosity samples, we recommend Set 1 and for highly viscous and aggressive samples Set 2. While the sample is added, you can check for air bubbles by looking through the inspection glass. A suitable medium is injected for the cleaning until all sample residues have been dissolved and removed. The drying unit will then remove all liquid residues.

  • Sample supply and cleaning via syringe
  • Intuitive operation via touch-screen display
  • Easy, menu-driven adjustment
  • Pre-defined scales such as density, relative density, brix, alcohol and sulfuric acid concentration
  • Compliance with GMP/GLP, 21 CFR Part 11, pharmakopoeias (USP, BP, JP, Ph. Eur.), FDA, ISO, HACCP, OIML, ASTM, ICUMSA, NIST

Typical applications

A comprehensive quality assurance covering the entire production process is a must in any industrial sector. Density measurements are frequently used for this purpose, especially in the pharmaceutical, chemical, petrochemical as well as the food and beverage industry. They allow the manufacturer to analyse raw materials, semi-finished and finished products as well as the manufacturing steps in terms of a number of factors.

Density can be used to identify substances, to determine their quality or purity and to measure their concentration in binary or quasi-binary mixtures. Substance conversions and reaction dynamics can also be inferred from it. In combination with other methods such as refractometry that measures the refractive index of substances, the density measurement allows you to make precise statements about the quality of each step of the production process.

Density meters based on the oscillating U-tube method are characterized by high ease of use and are therefore used in particular for continuous density measurement and for rapid routine measurements with larger numbers of samples in laboratories.

More information about typical applications of density meters can be found in our campus article about density meters – fields of application

Compliance with national and international standards

A big advantage of normalized measurements is the comparability of the measurement results. Many international standards and guidelines recommend density meters for specific measuring tasks as suitable measuring instruments. The requirements for the measuring devices, such as measurement accuracy and measuring temperature, are dependent on the respective measuring task and defined in each standard. We have compiled for you an overview of all known standards and guidelines, which refer to the density measurement.

  • American Society for Testing and Materials (ASTM)
  • Hazard Analysis and Critical Control Points (HACCP)
  • International Organization of Legal Metrology (OIML)
  • National Institute of Standards and Technology (NIST)
  • International Commission on Uniform Methods of Sugar Analysis (ICUMSA)
  • European Pharmacopoeia (Ph. Eur.)
  • United States Pharmacopeia (USP)

Density Meters Standards (PDF, 78 KB

Any questions?
Get in contact with us, we will assist you in choosing the right density meter for you.

Density measurement of liquids, emulsions and pastes

Thanks to the chemical-resistant sample carrying parts made of borosilicate glass and PTFE, our density meters are suitable for almost all liquids, emulsions and pastes.

The sample is administered manually via syringe, semi-automatically via peristaltic pump or fully automatically via autosampler. High-viscosity samples are usually supplied by syringe, low to light viscous can also be supplied by means of peristaltic pump and autosampler. For every method of operation and every kind of sample, we have suitable density meter sets available that contain all the necessary accessories, from tubing sets to filling spouts and adapters to splash guards for the manual supply of aggressive substances.

In the following table we have compiled an overview of typical samples from our customers. Most samples were measured with an measurement accuracy of 3 decimal places and at 20 ° C, deviations are indicated. The listed readings are not specifications.

Samples Density (PDF, 110 KB)

Any questions?
Get in contact with us, we will gladly assist you in selecting a suitable density meter for you.

Model specifications

 DS7700DS7800
MEASUREMENT DATA 
SCALESDensity [g/cm³]
Relative Density
Brix [%Brix]
Concentration of alcohol [vol%]
Concentration of sulphuric acid [wt%]
User-defined
MEASUREMENT RANGE0–3 g/cm³
MEASUREMENT ACCURACY±0.001 g/cm³±0.0001 g/cm³
MEASUREMENT PERIODTypically: 1–3 minutes including temperature control
SAMPLE VOLUME IN CASE OF MANUAL INJECTION0.9 ml
AMBIENT TEMPERATURE10–40 °C
TEMPERATURE RANGE10–40 °C
ACCURACY OF TEMPERATURE CONTROL±0.02 °C
METHODSA practically unlimited number of methods can be set
ADJUSTMENTAutomatic (menu-driven), with dried air and distilled water
CONTENT SET 1 
DS7700-1
DS7800-1
Drying unit with 2/2-way valve
Tygon tube set for use with Luer syringe: Air tube, waste tube, tube connections Luer
Luer nozzle
Luer syringes
PE waste container with lid
CONTENT SET 2 
DS7700-2
DS7800-2
Drying unit with 2/2-way valve
Tygon tube set for use with Luer syringe: Air tube, waste tube, tube connections Luer
PTFE tube set for use with Luer syringe: Drain tube, waste tube, PEEK hollow screws
PTFE splash guard
Adaptor Olive/UNFa for sample discharge through Tygon tube via UNF nozzle
Luer, UNF nozzle
Luer syringes
PE waste container with lid
DR6000-TDR6100-TDR6200-TDR6300-T
MEASUREMENT DATA 
SCALESRefractive index (nD)
Concentration of sucrose, glucose, fructose and invert sugar [%Brix],
User-defined
MEASUREMENT RANGEnD 1.3200–1.5800
0–95 %Brix
nD 1.3200–1.7000
0–95 %Brix
nD 1.32000–1.58000
0–95 %Brix
nD 1.32000–1.70000
0–95 %Brix
MEASUREMENT ACCURACYnD ±0.0001
±0.1 %Brix
nD ±0.00002
±0.02 %Brix
RESOLUTIONnD 0.0001
0.1 %Brix
nD 0.00001
0.01 %Brix
MEASUREMENT PERIODapprox. 4 s
MEASUREMENT PRISMSapphire
LIGHT SOURCELED
WAVELENGTH589 nm
TEMPERATURE DATA 
TEMPERATURE CONTROLWith integrated Peltier temperature control
TEMP. CONTROL RANGE10–80 °C
ACCURACY OF TEMP. CONTROL±0.1 °C
TEMPERATURE COMPENSATIONCan be activated
(ICUMSA or freely definable)
TEMPERATURE MEASUREMENTWith integrated temperature sensor Pt100
TEMPERATURE MEASUREMENT RANGE10–80 °C
TEMP. MEASUREMENT ACCURACY±0.1 °C